The Bruker SkyScan 1272 is a nondestructive micro-CT system for high-resolution three-dimensional imaging of small objects. The instrument is comparable to medical CT scans but on a much smaller scale. Micro-CT utilizes X-rays to create detailed cross-sectional images of an object, providing valuable insights into its internal structure and composition. This technology is used in applications across various fields, including materials science, biology, engineering, and archaeology. By capturing a series of X-ray images from different angles, Micro-CT reconstructs a 3D model of the sample, enabling researchers to visualize and analyze its internal features. The technique is particularly useful for examining objects with complex or delicate structures that cannot be easily dissected or sectioned.
Samples can be up to 75 mm in diameter. The system optimizes the X-ray energy and energy filtering using a new maintenance-free X-ray source and automatic 6-position filter changer. The system also includes a built-in precision micro-positioning stage. An optional 16-position automatic sample changer that allows for multi-sample data collection (maximum sample diameter of 15 mm).
For the fastest data collection times on this instrument (20-40 minutes), samples less than 10 mm in diameter are ideal. For materials larger or more X-ray dense (rock or minerals), data collection time can vary from 3-10 hours, depending on image quality settings. If you have any questions about the instrument or would like a consultation about your specific sample(s), please email [email protected].
Technical specifications for the SkyScan 1272
X-ray source: 20–100 kV, 10 W, <5 µm spot size at 4 W
X-ray detector: 11 Mp, 14-bit cooled CCD fiber-optically coupled to a scintillator
Maximum object size: 75 mm in diameter using offset scan (27 mm in fast single scan)
Detail detectability: 0.45 µm (11 Mp) at highest magnification
Optional stages: cooling, heating, compression/tension (not available in MATFab)
Radiation safety: <1 µSv/h at 10 cm from the instrument surface
This instrument is located in IATL 196 and is managed by MATFab staff.
Use BookitLab to reserve this instrument