Anton Paar (Quantachrome) Autosorb iQ
The Autosorb iQ is a volumetric gas sorption analyzer. A degassed solid sample is held at a fixed temperature while small doses of gas are admitted, and the amount adsorbed at each pressure gives the adsorption isotherm. From the isotherm the software calculates BET surface area, pore volume and pore size distribution (BET, BJH and DFT methods).
Use the Autosorb iQ when routine nitrogen BET on the Nova 4200e is not enough: micropores, very low surface areas (krypton), or CO2 and vapor uptake. The instrument is new to MATFab.
Quick Info Box
- Location: IATL 196
- Measurements: N2 at 77 K, Ar at 87 K, Kr at 87 K, CO2, vapor sorption
- Results: BET surface area, pore volume, pore size distribution
- Samples: dry solids that tolerate vacuum and heating
- Manager: Michael Sinnwell
- Contact: [email protected]
Specifications
- Volumetric (manometric) gas sorption
- Krypton physisorption for very low surface area samples and thin films
- Analysis methods include BET, BJH and DFT
Applications
- Zeolites, activated carbons and metal-organic frameworks: micropore size and volume
- Catalysts, ceramics, battery and membrane materials: surface area and porosity
- Low surface area powders and films: krypton BET
- Gas and vapor uptake: CO2 and vapor isotherms
Reservation & Access
- How to Access:
Contact Michael Sinnwell directly, or email [email protected] to discuss your project and samples. - Training Requirements:
Required prior to independent use. Training details provided upon request.
Use BookitLab to reserve this instrument
Anton Paar Autosorb gas sorption analyzer (manufacturer's product image); the MATFab instrument may look different. Image: Anton Paar.