ICP-OES

ICP Optical Emission Spectrometry

Agilent (Varian) 725 ICP-OES. Major and minor elements in solutions, measured simultaneously over 167 to 785 nm. Suited to higher concentrations and to solutions with more dissolved solids than ICP-MS accepts.

Rate: $50 / hour (academic). All rates

Hazardous materials: Highly acidic or toxic samples should be disclosed in advance.

Specialist: Linhan Li, Core Facility Research Specialist

Email Li · Book on BookitLab

ICP-MS

Agilent 7800 Quadrupole ICP-MS

Agilent 7800 quadrupole ICP-MS. Multi-element analysis of dissolved samples from 6Li to 238U, down to parts per trillion. Samples must be clear, dilute solutions; solids are digested first. Also supports single-particle ICP-MS (SP-ICP-MS) for nanoparticles.

Rate: $50 / hour (academic). All rates

Hazardous materials: Radioactive or highly toxic samples require advance notice and special handling approval.

Specialist: Linhan Li, Core Facility Research Specialist

Email Li · Book on BookitLab

Compare the instruments

PropertyICP Optical Emission SpectrometryAgilent 7800 Quadrupole ICP-MS
SummaryAgilent (Varian) 725 ICP-OES. Major and minor elements in solutions, measured simultaneously over 167 to 785 nm. Suited to higher concentrations and to solutions with more dissolved solids than ICP-MS accepts.Agilent 7800 quadrupole ICP-MS. Multi-element analysis of dissolved samples from 6Li to 238U, down to parts per trillion. Samples must be clear, dilute solutions; solids are digested first. Also supports single-particle ICP-MS (SP-ICP-MS) for nanoparticles.
Choose it whenChoose the ICP-OES for major and minor elements (ppm to percent levels), or for solutions too concentrated or too salty for ICP-MS.Choose the ICP-MS for routine trace and ultra-trace analysis of many dilute solutions: lower detection limits than the ICP-OES, and a lower rate than the HR-ICP-MS.
StrengthsMeasures all wavelengths at the same time, 167 to 785 nm
Several wavelengths per element extend the dynamic range and avoid spectral interferences
Tolerates more dissolved solids than ICP-MS
Detects parts per trillion or better
Collision/reaction cell removes argon-based interferences (for example on As and Se)
Autosampler holds more than 200 samples
Single-particle ICP-MS for nanoparticles
LimitationsDetection limits much higher than ICP-MS; not for ultra-trace work
Samples must be dissolved
Dissolved solids above 0.1 wt.% cannot be analyzed; samples must be diluted
Lower mass resolution than the HR-ICP-MS for difficult interferences
Solids must be digested first
Rate$50 / hour (academic)$50 / hour (academic)
ApplicationsMetallurgy & alloys
Mining & ores
Ceramics & glasses
Environmental
Quality control
Pharmaceutical
Materials science
Biological samples
Pharmaceutical analysis
Food science
Environmental testing
Nanoparticle characterization (SP-ICP-MS)
Sample typesDissolved solutions
Digested solids
Aqueous samples
Dissolved solutions
Digested samples
Liquids
Nanoparticle suspensions
Sample prepSamples dissolved in aqueous acid (HNO3 or HCl)
Higher tolerance for TDS than ICP-MS (up to 2%)
Minimum volume: 5 mL per sample
Concentration range: ppm to % level (dilute if too concentrated)
Samples must be dissolved in aqueous solution (2-5% HNO3 recommended)
No HF without prior approval from staff
TDS must be below 0.1% (dilute if needed)
Minimum volume: 3 mL per sample
Provide estimated concentration range
Include at least one blank and reference standard
For SP-ICP-MS: provide information on expected particle size range and concentration
ConstraintsHazardous: Highly acidic or toxic samples should be disclosed in advance.Hazardous: Radioactive or highly toxic samples require advance notice and special handling approval.
Note MATFab runs two ICP-MS instruments: this Agilent 7800 quadrupole for routine trace work, and the Thermo Fisher HR-ICP-MS for ultra-trace and isotope-ratio work. Contact staff if you are unsure which fits your project.

Still not sure which instrument to use?

Staff will help you choose the best fit for your project. Ask for guidance (opens an email to staff, already filled in).